| ABSTRACT |
|---|
|
|
|---|
Methods: Sinclair swine with spontaneous melanoma metastatic to regional lymph nodes were used because of their similarity to human melanoma. Organic lymphatic tracers tested included FDA-approved indocyanine green adsorbed non-covalently to human serum albumin (HSA), and NIR fluorophore CW800 conjugated covalently to HSA (HSA800). The inorganic/organic hybrid tracer tested was type II NIR quantum dots with an anionic coating. Primary tumors received four peri-tumoral injections of each tracer, with a fluorophore dose of 100 pmol to 1 nmol per injection. SLN mapping and image-guided resection were performed in real-time.
Results: Each of the 3 lymphatic tracers was injected into n = 4 separate primary melanomas in a total of 6 animals. All 12 injections resulted in identification of the SLN(s) and their associated lymphatic channels within 1 minute in 100% of cases, despite highly pigmented skin and black fur. Hydrodynamic diameter had a profound impact on tracer behavior in vivo.
Conclusions: This study completes the pre-clinical development of NIR fluorescence-guided SLN mapping and provides insight into imaging system optimization and tracer choice for future human clinical trials. The technology is likely to eliminate the need for radioactive and colored tracers, permits real-time image guidance throughout the procedure, and assists the pathologist in tissue analysis.
Key Words: Sentinel lymph node mapping Pre-clinical development Near-infrared light Near-infrared fluorescence Melanoma Sinclair pigs
| INTRODUCTION |
|---|
|
|
|---|
1 µm), the need for direct tissue contact, and the limited field-of-view (FoV) restrict its utility. Light, and specifically near-infrared (NIR) light in the biological window (700900 nm), can be exploited for intraoperative imaging guidance for four important reasons. First, it is invisible, so the look of the surgical field is not altered. Second, it is safe, since the fluence rate (i.e., amount of light) used is small and zoom lenses ensure that the instrument is far from the surgical field. Third, wavelengths in the 800 nm range penetrate relatively deeply into living tissue compared to visible light.1 Fourth, and most importantly, there has been tremendous recent effort in developing general-purpose NIR fluorophores that can be conjugated to targeting or other molecules, thus creating "contrast agents" matched to any desired surgical application (reviewed in 2).
Our laboratory has previously described NIR fluorescence imaging systems designed for small animal3,4 and large animal5,6 surgery. The latter is termed the FLARE system, for Fluorescence- Assisted Resection and Exploration. Since our ultimate goal is to translate this technology to the clinic, recent engineering efforts have begun to add hands-free operation and cardiac and respiratory gating,7 as well as training tools for surgeons in NIR fluorescence-guided surgery.8 In parallel, we have developed robust methods for the conjugation of heptamethine indocyanine NIR fluorophores to small molecules,9 peptides,9 and proteins.10 Such conjugates create specialized contrast agents that can be optimized in animal model systems.
Because of immediate clinical need, we have focused on the development of a family of lymphatic tracers optimized for sentinel lymph node (SLN) mapping and resection. Currently, SLN mapping utilizes a combination of radioactive tracer (typically Tc-99m based) and blue dye. Tc-99m-based tracers are used with pre-operative gamma camera imaging and/or an intraoperative handheld gamma probe to generally localize the SLN, and the blue dye to pinpoint the SLN. However, this system adds time to the procedure, requires the involvement of a nuclear medicine physician, exposes patient and caregivers to ionizing radiation, and requires contact (gamma probe) with tissue. The current procedure also has a steep learning curve, requiring up to 60 cases for technical proficiency in breast cancer.11 Until proficiency is attained, up to 16% of SLNs can be missed.12
In previous studies, we demonstrated the utility of intraoperative NIR fluorescence imaging for SLN mapping in skin,6,10 lung,13 pleural space,14,15 mammary tissue,6 esophagus,16 stomach,17 small intestine,10,17 and colon.17 Despite the use of large animals approaching the size of humans, these studies could be criticized for not employing a model system with spontaneous cancer metastatic to regional lymph nodes. We now utilize such a model to complete pre-clinical development of the technology and discuss the factors that will govern its translation to the clinic.
| MATERIALS AND METHODS |
|---|
|
|
|---|
NIR excitation light is provided by an array of 10 high-powered (1W) light emitting diodes (LEDs; L-760-66-60, Marubeni Epitex, New York, NY), which are filtered (HHQ750/50x, Chroma Technology, Brattleboro, VT) and focused on the surgical field. White light is provided by an array of four Luxeon III white light diodes (Lumileds, San Jose, CA) from which all NIR wavelengths are filtered out (E680SP, Chroma). The generated fluence rates for NIR excitation and white light range from 05 mW/cm2 and 01 mW/cm2, respectively.
Spatial resolution at a field-of-view of 20 x 15 cm is 625 µm, and at a field-of-view of 4 x 3 cm is 125 µm. After computer-controlled (LabVIEW) camera acquisition via custom LabVIEW (National Instruments, Austin, TX) software, anatomic (white light) and functional (NIR fluorescence light) images can be displayed separately and merged. To create a single image that displays both anatomy (color video) and function (NIR fluorescence), the NIR fluorescence image is pseudo-colored in lime green and overlaid with 100% transparency on top of the color video image of the same surgical field. All images are refreshed up to 15 times per second. The entire apparatus is suspended on an articulated arm over the surgical field, thus permitting non-invasive and nonintrusive imaging. Hands-free operation utilizing motorized zoom and focus lenses, a four-pedal foot-switch, and a multifunction data acquisition board were described in detail previously.7
Organic Lymphatic Tracers
Two different organic tracers that fluoresce at
800 nm were tested. Indocyanine green (ICG) is a 776 Da di-sulfonated small molecule first FDA approved in 1958 for indicator dilution studies; it is also a NIR fluorophore that has a moderate affinity binding site on human serum albumin (HSA). As we demonstrated previously, ICG adsorbed non-covalently to HSA (ICG:HSA) exhibits a 3.6-fold increase in quantum yield (i.e., brightness), and the 70 kDa complex can be used as a lymphatic tracer.10 HSA800 is a covalent conjugate of IRDyeTM 800CW NIR dye (CW800; LI-COR, Lincoln, NE; MW 962 Da) to HSA at a molar ratio of 3:1. Its synthesis and purification has been described previously.10 Final products were resuspended at a final concentration of 10 µM in phosphate-buffered saline, pH 7.4 (PBS) prior to injection.
Inorganic Lymphatic Tracer
Type II NIR fluorescent quantum dots (QDs) were synthesized with a hydrodynamic diameter of 15 nm, a maximal absorption cross-section, an 800 nm fluorescence emission, an aqueous quantum yield
9%, and a stable oligomeric phosphine coating as described previously.6,18,19 Final products were resuspended at a concentration of 1 µM in PBS prior to injection.
High-Resolution Gel-Filtration Analysis
Gel-filtration analysis with on-line absorbance (200870 nm) and fluorescence (2001100 nm) spectrometry was performed as described in detail previously.20 To measure the effect of serum on hydrodynamic diameter, tracers were diluted in either PBS or 100% calf serum prior to loading of a Supe-rose-6 10/300 GL gel-filtration column (Amersham Biosciences, Piscataway, NJ).
Large Animal Model of Spontaneous Melanoma
Animals were housed in an Association for Assessment and Accreditation of Laboratory Animal Care (AAALAC)-certified facility staffed by full time veterinarians and were used under supervision of an approved institutional protocol. Sinclair mini-swine of 812 weeks of age, and weighing 810 kg, were purchased from Sinclair Research Center, Inc. (Columbia, MO). They were housed at a pen temperature of 7580°F and fed S-9 Sinclair Swine Ration #33133P (Milbank Mills, Chillicothe, MO) and supplements of brown Karo corn syrup. Each animal had 24 spontaneous primary melanomas ranging in size from 12.5 cm, with most having either microscopic or macroscopic regional nodal metastases. General anesthesia was induced with 10 mg/kg of intramuscular Telazol (Fort Dodge Labs, Fort Dodge, IA). Animals were intubated with a 3 mm endotracheal tube and anesthesia was maintained with 1.52% isoflurane/balance O2 using an MRI-1 ventilator with an expansion valve set (CWE, Inc., Ardmore, PA).
Lymphatic Mapping and SLN Identification
A typical clinical SLN injection protocol was used. 0.4 cc of NIR fluorescent lymphatic tracer resuspended in PBS was taken up into a 1 cc plastic syringe fitted with a 27 gauge,
" needle. With bevel facing up, each primary melanoma was injected peri-tumorally, in a 4-quadrant pattern, with four 0.1 cc injections. No attempt was made to cannulate lymphatic channels; the needle was simply placed subcutaneously. After injection, the tumor was massaged gently with the thumb four times. Four primary tumors were injected with each lymphatic tracer. Injections, lymphatic tracking, and SLN resection were imaged in real-time, and images archived on the computer. Time from injection to SLN identification was measured using a time-stamp on the archived images. Post-resection inspection of the surgical field was performed using an extended integration time to ensure completeness.
Histological Analysis
Nodal tissue was cryo-sectioned at 6 µm onto Superfrost Plus slides, fixed in 80% acetone/20% water, and stained with hematoxylin and eosin (H&E).
| RESULTS |
|---|
|
|
|---|
|
|
Large Animal Model System of Spontaneous Melanoma
The Sinclair mini-swine model of spontaneous melanoma was developed in the early 1990s21 and has recently been used extensively by Goldberg and colleagues for the optimization of ultrasound-guided SLN mapping.22,23 At birth, 54% of animals already have primary melanomas, which increases to 85% at one year.23 Of importance is that over 70% of primary lesions result in metastases to regional lymph nodes, and spontaneous regression of tumors is uncommon prior to 3 months of age. Coupled with its remarkable histological similarity to human melanoma, this strain of Sinclair mini-swine is an ideal surgical model for the validation of new techniques for SLN mapping. Sinclair swine also represent a "worst case" scenario for optical imaging since they have highly pigmented skin (corresponding to types V-VI human skin) and black fur, which scatters light.
NIR Fluorescent Lymphatic Tracers
Three different lymphatic tracers developed by our laboratory6,10 were used in this study. The relative size (i.e., hydrodynamic diameter; HD) of the agents and their detailed physicochemical and optical properties are presented in Fig. 2A
. Organic tracers are based upon human serum albumin (HSA), a naturally occurring protein in high abundance in serum. HSA800 is a covalent conjugate of 3 molecules of fluorophore CW800 per molecule of HSA and has a high fluorescent yield. ICG:HSA is a formulation of FDA-approved indocyanine green (ICG) adsorbed noncovalently with HSA. It has a lower total fluorescent yield than HSA800 and NIR QDs, but is composed of two compounds already FDA-approved for other indications. Both of these organic tracers have a HD of
7 nm, permitting them to quickly enter lymphatics, and travel rapidly to the SLN; they will also flow through the SLN if given enough time (discussed below). NIR QDs are inorganic/organic hybrid molecules with extreme photo-stability and high fluorescent yield. They are also engineered to have a large size (
15 nm), which guarantees retention in the SLN.
|
20 nm, with a broadening of the shape distribution, and the appearance of very high molecular weight forms. These gel-filtration results help explain why, in eight separate studies,6,10,13,1517,20,24 NIR QDs are always retained at the first node encountered.
Real-Time NIR Fluorescent SLN Mapping
Each of the 3 lymphatic tracers was injected into n = 4 separate primary melanomas in a total of 6 animals. The entire procedure, including injection (Fig. 3
), could be followed in real-time, with the otherwise invisible NIR fluorescent light identifying all lymphatic channels draining from the primary tumor, and within 15 seconds after injection, identifying the SLN(s). Indeed, the SLN(s) was identified in 100% of injections with all three tracers, although as expected based on total fluorescence yield (Fig. 2A
), ICG:HSA exhibited slightly lower signal strength than HSA and NIR QDs.
|
Pathological Analysis
After tissue resection, the imaging system could be used to dissect nonnodal tissue away from the SLNs and to show the surgical pathologist the sinus entry point (Fig. 4A
). Tissue from this area was then analyzed histologically. As shown in Figure 4B
, both the proximal and distal SLNs were found to harbor metastatic tumor cells, while the nearby control LN had none present.
|
| DISCUSSION |
|---|
|
|
|---|
Since even NIR light is scattered in living tissue, the depth of light penetration might be expected to be a major obstacle to optical imaging in surgery; however, this doesnt seem to be a major issue. First, the surgeon can follow the lymphatic channels from the injection site towards the SLN, so the direction is always known. Second, because tissue background is so low, tracers can be detected below relatively thick (
0.51 cm) tissue6 and below up to 5 cm of lung.13 Of course, on the surface of the tissue, the appearance of the SLN will be "blurred" by scatter and appear larger than it really is (see, for example, Fig. 3
), but its exact location, and therefore the exact location of the surgical incision, will still be known. Finally, it will likely be possible to adapt newer optical techniques, such as frequency-domain photon migration (reviewed in 25,26), to our imaging system to improve detectability into the 24 cm range.
This study essentially completes pre-clinical development of NIR fluorescent SLN mapping and resection. Using large animal model systems, we have now demonstrated optimized protocols for SLN mapping of spontaneous melanoma, skin,6,10 lung,13 pleural space,14,15 mammary tissue,6 esophagus,16 stomach,17 small intestine,10,17 and colon.17 Using this technology, the need for radioactivity and blue dye is eliminated, and the SLN is typically identified within seconds.
Our studies also found important differences among lymphatic tracers; these differences have implications for the translation of this technology to the clinic. Organic tracers, like those from this study based on albumin, are advantageous because of their availability and biocompatibility. Inorganic/organic hybrid NIR QDs are brighter and remarkably stable to photobleaching, but have yet to undergo proper toxicological screening. From this study, it also appears that type II NIR QDs with a purely anionic coating bind serum proteins and shift their size to 20 nm and greater, whereas zwitterionic molecules with the same concentration of anions, do not. This is an important finding because the small agents, such as HSA800 and ICG:HSA, often identify lymphatic branches (see 16 for detailed description and Figure 3
of this study) and small SLNs, whereas we have never seen branching with NIR QDs and other large HD tracers. The importance of this point cannot be overstated. As shown in Figures 3
and 4
, the other (distal) SLN was extremely small (only 2 mm) yet harbored metastatic cells. For eventual use in humans, the goal must be to find all SLNs, regardless of their size and the size of the connecting lymphatic channels. Since flow in lymphatic channels is pulsatile with heartbeat and respiration, and channels are often collapsed, the smaller HD particles might have an advantage. Moreover, since the SLN is identified within seconds using NIR fluorescence imaging, the probability of tracer flow through the SLN to the next tier is minimized. At least in the case of the Sinclair melanoma model, SLNs from all 12 primary tumors studied harbored either micro- or macrometastases, suggesting an extremely low false-negative rate for all three lymphatic tracers.
The clinical translation of this technology for SLN mapping is imminent. At the present time, the ICG:HSA formulation offers the significant advantage that both components are already FDA approved for other indications, the amount of each component needed for SLN mapping is orders of magnitude lower than the doses of each used for these other indications, and total fluorescence yield in a serum environment is reasonably high. Within the organic class of SLN tracers, HSA800 would provide an additional 2.2-fold improvement in fluorescent yield, but HSA800, like ICG:HSA, would have a hydrodynamic diameter that permits passage through the SLN if too much time elapses after injection.10,14 Quantum dots, although still years away from clinical translation, provide the highest fluorescence yield, as well as a definable hydrodynamic diameter that ensures complete retention in the first lymph encountered.6,10,13,1517,20
Newer NIR fluorescent contrast agents, such as HSA800,10 are being developed at a relatively high rate, and the future promises NIR fluorescent contrast agent targeted to a variety of cancers.27 Indeed, tumor-specific targeting should permit real-time assessment of tumor margins and identification of occult metastases within the surgical field. The technology can also help the surgeon to avoid nerves, blood vessels, etc. during cancer resection. The availability of tissue and disease-targeted NIR fluorophores will likely have a lag of 35 years due to a variety of regulatory factors (reviewed in 28). Nevertheless, ICG:HSA, and ICG alone,5,29,30 will undoubtedly have many clinical applications and will hopefully fill the void.
It should be emphasized that the NIR fluorescence imaging system described in this study might eventually play a dual role in patient management. It not only provides the surgeon with realtime image guidance during any oncologic procedure, but also assists the surgical pathologist. First, the specimen can be trimmed under image guidance. Second, the particular area of interest, e.g., sinus entry point of the SLN, can be found and sectioned at high-resolution to find micrometastases. Finally, since NIR fluorophores withstand frozen sectioning,6,16 a NIR fluorescence microscope31 can be used to correlate the resected target with at least three other fluorescence channels of biologic markers.
| ACKNOWLEDGMENTS |
|---|
Received for publication July 21, 2006. Accepted for publication July 21, 2006.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Ogawa, N. Kosaka, P. L. Choyke, and H. Kobayashi In vivo Molecular Imaging of Cancer with a Quenching Near-Infrared Fluorescent Probe Using Conjugates of Monoclonal Antibodies and Indocyanine Green Cancer Res., February 15, 2009; 69(4): 1268 - 1272. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. V. Frangioni New Technologies for Human Cancer Imaging J. Clin. Oncol., August 20, 2008; 26(24): 4012 - 4021. [Abstract] [Full Text] [PDF] |
||||
![]() |